R Krauss, M Morales, C Coffey… - Medical Physics, 2011 - link.aip.org Purpose: To examine the accuracy of different methods of converting percentage depth dose (PDD) curves to tissue phantom ratios (TPR) for small cones by comparing them with directly measured TPR values.Methods: For a 6 MV beam, a detector remained fixed in water at ...
G Wu, G Courlas, D Lewis… - Medical Physics, 2011 - link.aip.org Purpose: To validate the use of Varian Golden Beam Data PercentageDepthDose (PDD) information for the Eclipse AAA algorithm for newly installed Varian Linear Accelerators Methods: Comparison was made between measured data and Varian Golden Beam Data which is ...
W Gebreamlak, D Tedeschi… - Medical Physics, 2011 - link.aip.org Purpose: To calculate the percentagedepthdose of any irregular shape electron block using the lateral build‐up‐ratio method.Methods: Percentagedepthdose (PDD) curves were measured using 6, 9, 12, and 15 MeV electron beam energies for applicator cone sizes of 6×6 ...
B Arjomandy, R Tailor, L Zhao… - Medical Physics, 2011 - link.aip.org ... For each beam, film response measured over the dose range 0.6–6.2 Gy, was employed for OD‐to—dose conversion. Results: for all mega‐voltage photon & electron beams, percentage depth‐doses measured with EBT2 showed an excellent agreement (<1.5%) with those ...
LIR Garcia… - Medical Physics, 2011 - link.aip.org ... Phys. 36(11), 5261–5291 (2009). LJ van Battum, M. Essers, and PR Storchi, “Conversion of measured percentagedepthdose to tissue maximum ratio in stereotactic radiotherapy,” Phys. Med. Biol. 47, 3289–3300 (2002). EB ... Related articles